Review ArticleOpen Access, Volume 4 Issue 1

A review of the performance of rapid diagnostic tests for the NS1 protein in diagnosing dengue infection

Zapata de la Cruz Melina, MSc1; Falcón Lezama Jorge A, PhD2; Quiroz Gómez Sergio, PhD3; Guzmán Priego Crystell G, PhD4*

1Cardiometabolism Laboratory, Academic Division of Health Sciences, Universidad Juárez Autónoma de Tabasco, Villahermosa, Tabasco, Mexico.

2Carlos Slim Foundation, Mexico City, Mexico.

3Academic Division of Health Sciences, Universidad Juárez Autónoma de Tabasco, Villahermosa, Tabasco, Mexico.

4Cardiometabolism Laboratory, Academic Division of Health Sciences, Universidad Juárez Autónoma de Tabasco, Villahermosa, Tabasco, Mexico.

*Corresponding author: Guzmán Priego Crystell G

Cardiometabolism Laboratory, Academic Division of Health Sciences, Universidad Juárez Autónoma de Tabasco, Villahermosa, Tabasco, Mexico.
Tel: +52 1 993 242 8993;

Email: crystell_guzman@hotmail.com

Received : Dec 02, 2025       Accepted : Feb 11, 2026       Published : Feb 18, 2026

Epidemiology & Public Health - www.jpublichealth.org

Copyright: Crystell GP © All rights are reserved

Citation: Melina ZC, Jorge FL, Sergio QG, Crystell GP. A review of the performance of rapid diagnostic tests for the NS1 protein in diagnosing dengue infection. Epidemiol Public Health. 2026; 4(1): 1088.

Abstract

Introduction: The dengue virus is composed of structural proteins (envelope, membrane, and capsid) and non structural proteins (NS1, NS2A, NS2B, NS3, NS4A, NS4B, and NS5). The NS1 protein is the most relevant, as its hexameric form is released into the blood and can be detected in this medium, in serum and plasma. Rapid diagnostic tests are presented as an alternative to detecting dengue infection through this protein due to their accessible format.

Objectives: To establish a review of the performance of three commercial rapid diagnostic tests available in Mexico, based on the detection of the NS1 protein for the diagnosis of dengue through sensitivity and specificity values.

Methods: Twelve retrospective diagnostic research stud ies were reviewed from scientific databases that included the SD Bioline Dengue Duo ABBOTT, CERTUM Rapid Dengue Combo Test Cassette, and Standard F Dengue NS1 Ag FIA SD Biosensor tests and reported their sensitivity and specificity values.

Conclusions: Most of the articles reviewed show good performance, with the highest being the Standard F Dengue NS1 Ag FIA SD Biosensor. However, factors such as primary and secondary infections, serotypes, and cross-reactions also play an important role in the performance of these tests.

Keywords: Dengue virus; Rapid diagnostic test; NS1 protein; Lateral flow assay; Infectious disease

Introduction

The dengue virus is classified as an arbovirus with four dis tinct serotypes: DENV 1-4 throughout tropical and subtropical regions [1-3]. Rapid Diagnostic Tests (RDT) are tools for improv ing the diagnosis of dengue infections in remote or low-resource areas, can be used with basic training and results available in less than 20 minutes. However, there are conflicting reports re garding the sensitivity and specificity of these tests [4,5]. This review aims to provide an overview of current international and national reports on the sensitivity and specificity of three rapid tests for the NS1 protein that are commercially available and easily accessible.

Materials and methods

This review considered retrospective diagnostic accuracy research articles published in English and Spanish. The articles were retrieved from scientific databases such as PubMed, Re search Gate, Web of Science and Google Scholar between March 2024 and May 2025, as described in Figure 1. Addition ally, articles published within the last five years were given pri ority. Among the inclusion criteria, the use of at least one of the following tests had to be reported in the literature consulted: SD Bioline Dengue Duo ABBOTT (11FK46), CERTUM Rapid Den gue Combo Test Cassette (IDEC-425), and Standard F Dengue NS1 Ag FIA SD Biosensor (10DEN10D), which examine the NS1 antigen and, with the exception of the latter test, also assess IgM and IgG. The above rapid tests were considered because they are commercially available and easily accessible in Mexico.

The following search terms were used to search for information:

1. RDT dengue

2. SD Dengue Duo

3. Rapid Diagnostic Test dengue

4. SD Bioline Dengue Duo

5. Standard F NS1 Ag

6. (RDT)(dengue)

The evaluated literature had to include at least one of the following aspects in the study population: a history of dengue infection, location in an endemic area for the vector, or pre sentation of symptoms characteristic of the infection, such as high fever. In addition, a healthy control group or a group with another type of infection had to be included in the studies. An other criterion for the inclusion of articles was that a reference standard had to be present for at least each of the antigens and antibodies evaluated. In addition, it was necessary for them to include the sensitivity and specificity values for the NS1 protein. The values for IgM/IgG could be included or not included.

Those articles without a validated reference standard or sensitivity and specificity values for NS1 were excluded. No re strictions were placed on the age of the study population, the biological material used (serum, plasma, or blood), the identifi cation of primary or secondary infection, or the onset of symp toms for each item.

Results

The identification of thirty retrospective diagnostic accuracy research articles was the result of the initial research. Priority was given to the most recent articles published within the last five years, thus excluding seventeen articles published between 2010 and 2019. One more article was discarded because it did not report the NS1 specificity value. After applying the inclusion and exclusion criteria, only twelve articles were included in this review The most widely used rapid test during the review of the articles was the SD Bioline Dengue Duo ABBOTT, which account ed for 7 (58.3%) of the total number of studies. The Standard F Dengue NS1 Ag FIA SD Biosensor was the second most popular, with four (33.3%) articles, and finally, the CERTUM Rapid Den gue Test Combo Cassette had one article (8.3%). It should be mentioned that, to date, the cited study is the only one in which the performance of this last test has been reported.

A total of 12 articles were developed in endemic regions, ex cept for the work of Matusali et al. [6] carried out their work in a non-endemic area (Italy) but with samples from people with a history of international travel. All studies described in clusion criteria regarding age ranges for the febrile population with a minimum age of >9 months and no maximum age. Gen der was not considered an exclusion criterion for any article, as the sample population was selected based on febrile samples suspected of dengue infection. For this same reason, research ers conducted epidemiological surveillance studies or test per formance studies in the study area during past epidemic out breaks. As for the control groups, samples from individuals with fever due to other viral or bacterial infections were included in eleven articles, and just one article also included blood donors and healthy patients, further favoring the evaluation of specific ity. The biological material most used for both groups was se rum (8), followed by plasma (3) and finally blood (2). The most widely used reference standard for identifying the dengue virus and its serotypes was RT-PCR. All the articles presented report its use for comparison with the NS1 protein, although other techniques were also used, such as in the study by Yow et al. [7], which used viral isolation and an immunofluorescence test to identify the serotype. For serological tests, four articles used ELISA for IgM as the reference standard, while three articles used IgG alone. Another technique used for serological testing was the plaque reduction neutralization technique, also consid ered a reference standard, as described by Zammarchi et al. [8].

The SD Bioline Dengue Duo Abbott rapid diagnostic test and the CERTUM Rapid Dengue Test Combo Cassette identify the NS1 protein, as well as IgM and IgG. In contrast, the Standard F Dengue NS1 Ag FIA SD Biosensor detects only NS1. The per formance of the tests for NS1 alone was ranked in a sensitiv ity range of 38.6%–90% and a specificity range of 87.7%–100%. Between each test, the best sensitivity performance of SD Bio line Dengue Duo was 90% compared to the lowest, which was 38.6%. In terms of specificity, the range was between 90% and 100%, while the sensitivity range of Standard F Dengue NS1 Ag FIA was 79.11%–100% and its specificity was 87.5%–100%. Overall, the Standard F Dengue NS1 Ag FIA SD Biosensor per formed best in terms of sensitivity for NS1. It also had the best specificity values, along with the SD Bioline Dengue Duo AB BOTT.

A total of four articles evaluated primary infection with a ref erence standard method for IgM and the reported performance range was 4.4%–83.3% for the SD Bioline Dengue Duo ABBOTT. The reports showed conflicting results when evaluated with NS1, with some indicating benefits and others showing no clear advantages. Liu et al. [9] report that adding IgM or IgG increases the rapid test’s sensitivity in identifying secondary infections. This finding is supported by Yow et al. [7] and Chong et al. [10], who also found better results when IgM was included with NS1 than when NS1 was used alone. However, for the work of Kikuti et al. [11], the addition of IgM to the NS1 assessment is not suf ficient for an increase in sensitivity values.

Regarding the evaluation of secondary IgG infection, sensi tivity was reported in only two articles, with the best perfor mance being that of Santoso et al. [12] at 60%. However, the result may be influenced by the large difference in the number of samples that were analyzed.

In general, the serotypes of the dengue virus are evaluated in all articles, except for Zapata et al. [13] and Zammarchi et al. [8]. All articles identify the four serotypes except Kikuti et al. [11], who only described DENV-1, DENV-3, and DENV-4.

Additionally, five articles evaluated cross-reactivity with oth er flaviviruses, viruses from other families, and bacteria. Two articles evaluated these characteristics with SD Bioline Dengue Duo ABBOTT test. The study by Santoso et al. [12] reports that this test does not cross-react with any of the evaluated infec tions, achieving 100% specificity. In contrast, the study by Yow et al. [7] it was evaluated against samples of early and convales cent Zika infection, as well as samples of Chikungunya, and the test did not react with any of them. Two other articles evalu ated the cross-reaction of the Standard F Dengue NS1 Ag FIA SD Biosensor. First, in the study by Zammarchi et al. [8], of the total number of samples negative for dengue, cross-reaction occurred for the Toscana virus, rheumatoid factor, and Epstein Barr virus, with only one case reported for each condition. Sec ond, Zuroidah et al. [14] study reports 100% specificity, indi cating that there was no cross-reaction with other infections. Information on the overall performance of rapid diagnostic tests is summarized in Table 1.

Table 1: Performance of three rapid diagnostic tests for the NS1 protein of the dengue virus commercially available in Mexico.
Test Author Sample size Patient characteristics Sample type Referenced method Sensitivity Specificity
SD BIOLINE dengue duo Kikuti M et al. [11] Dengue cases: 246
Controls: 254
Suspected dengue, >5 y/o, 2–3 onset of feverControls, >5 y/o, blood donors, x/50 Serum ELISA-NS1
ELISA-IgM
RT-PCR
NS1: 38.8%
IgM: 13.83%
NS1/IgM: 46.8%
NS1: 89%
IgM: 94.9%
NS1/IgM: 88–96%
Santoso et al. [12] Dengue cases: 50
Controls: 49
Suspected dengue, <4 days and >4 days onset of fever Controls with febrile illnesses (malaria, typhoid fever and bacterial infections) Serum Panbio Dengue Duo IgM/IgG
RT-PCR
NS1: 73%
IgM/IgG: 60%
NS1: 100%
Liu et al. [9] Dengue cases: 13
Controls: 37
Suspected dengue, >8 y/o, 0–6 onset of fever Serum RT-PCR NS1: 89.7%
NS1 to IgM: 95.6%
NS1/IgM/IgG: 97.1%
NS1: 91.9%
NS1 to IgM: 82.9%
NS1/IgM/IgG: 86.5%
Prabowo et al. [15] Dengue cases: 99
Controls: 102
Suspected dengue, 1–60 y/o, <5 days and ≥6 days onset of fever Serum RT-PCR NS1: 87.88% NS1: 90%
Yuw et al. [7] Dengue cases: 108
Controls: 30
Suspected dengue, 4–11 days onset of fever Serum RT-PCR
ELISA NS1
ELISA IgM/IgG
NS1: 68.5%
NS1 + ELISA IgM/IgG: 97.2%
NS1: 100%
Chong et al. [10] Dengue cases: 227
Controls: 226
Suspected dengue, <5 days and >5 days onset of fever Capillary and venous blood RT-PCR
Capture IgM/IgG
ELISA
Panbio ELISA primary infection
NS1: 52.4%
IgM: 45.4%
IgG: 27.8%
NS1 to IgM: 75.8%
NS1: 97.7%
IgM: 98.5%
IgG: 88.9%
NS1 to IgM: 96.6%
Vickers et al. [16] Dengue cases: 309
Controls: 30
Suspected dengue Serum ELISA antigen NS1
ELISA IgM/IgG
NS1: 90%
IgM: 49.3%
IgG: 39.1%
NS1/IgM/IgG: 98.9%
NS1: 99.2%
IgM: 95.9%
IgG: 100%
NS1/IgM/IgG: 100%
Cassette combo prueba rápida Zapata et al. [13] Dengue cases: 45
Controls: 79
Suspected dengue, 2–7 days onset of fever
Standard F Zammarchi et al. [8] Dengue cases: 21
Controls: 36
Suspected dengue, <8 days onset of fever Controls with febrile illnesses (Zika, West Nile virus, Toscana virus, Epstein-Barr virus) Serum ELISA-DENV NS1 Ag
RT-PCR
ELISA IgM/IgG
NS1: 90% NS1: 87.5%
Zuroiadah et al. [14] Dengue cases: 50
Controls: 30
Suspected dengue, 3–7 days onset of fever Controls with febrile illnesses (typhoid fever, malaria, hepatitis B, hepatitis A, and leptospirosis) Serum ELISA-DENV NS1 Ag
RT-PCR
NS1: 82% NS1: 100%
Ruchusatsawat et al. [17] Dengue cases: 158
Controls: 246
Suspected dengue, 2–4 days onset of fever Controls with febrile illnesses (Chikv, bacterial infections, others) Plasma RT-PCR
ELISA IgM and ELISA IgG
NS1: 79.11% NS1: 92.28%
Matsuii et al. [6] Dengue cases: 66 Suspected dengue with history of international travel to endemic areas Plasma
Serum
Blood
Real time RT-PCR NS1: 84.6% NS1: 100%

Discussion

Although the performance of rapid tests depends largely on the test itself, the result may be influenced by other factors such as sample collection time, viremia, serotype, and even an tibodies responding to infection.

The best performance reported among the tests corresponds to Standard F, and it is important to note that although all three tests are lateral flow tests, detection in the aforementioned test is performed using fluorochromes or fluorescence-labeled par ticles that are detected by signal-sensitive equipment, unlike the other two tests, which use colloidal gold particles, among others, and show the test lines with the naked eye [18]. This difference could be significant, as the results are subject to the interpretation of each evaluator and biases could be intro duced, as has been reported in other studies where a decision is reached by up to three evaluators [10].

The timing of sample collection also influences the results, as researchers have reported that during a primary infection, the NS1 protein remains in serum on average from the onset of symptoms until the sixth day, although some cases have also been found to remain until the ninth day [19,20]. Based on this reasoning, most of the articles reviewed sampled before the sixth day and obtained sensitivity values above 73%. Neverthe less, there are investigations such as that by Yow et al. [7] that reported reduced sensitivity values even though some of their samples are within the acceptable range. The beneficial effect on performance might be due to the specific proportion of samples collected before the sixth day. In the study mentioned, the proportion was 41/67 samples obtained before and after six days, respectively. As the majority are acquired after the seventh day, when the amount of NS1 in the blood starts to de crease [21], it is less probable that the examination will identify it, an outcome that is strengthened by the examination’s high sensitivity to IgM, which rises after the fifth day [22].

On the other hand, Kikuti et al. [11] and Zapata et al. [13] re ported lower sensitivity values despite collecting samples with in the recommended range. Both studies were conducted in en demic areas constantly exposed to dengue infections, where it has been reported that in secondary infections, the NS1 protein decreases compared to a primary infection [23] due to an IgG response that reduces the viral load and, therefore, the con centration of soluble NS1. Also, it forms an NS1-IgG immune complex that blocks the epitopes of the NS1 protein, preventing them from being detected by the antibodies attached to diag nostic tests [24,25].

Serotypes also play a role in determining the sensitivity of rapid diagnostic tests. In their study, Liu et al. reports that the SD Dengue Duo test has higher sensitivity for detecting DENV-1 than DENV 2-4, a result has also been documented with other rapid tests, which report sensitivity values of 92.2% and 50% for DENV-1 and DENV-2, respectively [4,26]. This suggests that the sensitivity may have been affected by a wider distribution of DENV-2 samples.

Another factor that could also be related to the performance of rapid tests is cross-reactivity. However, in the studies evalu ated, this result was minimal, as in the case of SD Bioline, which was compared in one study against cell cultures infected with variants of the Zika virus (Uganda, Thailand, and Puerto Rico) and concluded that under natural conditions these viruses would not express false positives since the viral titers are lower than those detected by rapid tests for NS1 of the Dengue virus [27]. Another type of rapid diagnostic test known as TKK2, eval uated in Bangladesh, has shown the same condition. This test did not cross-react with samples infected with the Zika virus [28], and it did not show any reaction when evaluated against RT-PCR [29]. A compilation of studies and their results in SD Bio line and Standard F indicate that rapid diagnostic tests based on the dengue NS1 protein are highly specific compared to other flaviviruses [30].

Conclusion

Rapid diagnostic tests facilitate the diagnosis of dengue in fection and play a particularly important role in endemic areas where resources are limited, and even more so in certain emer gency situations. Although the results were inconsistent in this review, most of the studies showed good results in detecting the NS1 protein and even against other infections. To ensure optimal performance, it is essential that the use of these rapid tests adhere to the guidelines provided by the suppliers, consid ering prior work, and the conditions of use. It is also important to mention that the constant evaluation of the performance of rapid diagnostic tests allows for the development of this tech nology to have a greater and more accurate ability to detect dengue infection.

References

  1. Roy SK, Bhattacharjee S. Dengue virus: Epidemiology, biology, and disease aetiology. Can J Microbiol. 2021; 67: 687–702.
  2. Bhatt P, Sabeena SP, Varma M, Arunkumar G. Current understanding of the pathogenesis of dengue virus infection. Curr Microbiol. 2021; 78: 17–32.
  3. Tiga-Loza DC, Martínez-Vega RA, Undurraga EA, Tschampl CA, Shepard DS, Ramos-Castañeda J. Persistence of symptoms in dengue patients: A clinical cohort study. Trans R Soc Trop Med Hyg. 2020; 114: 355–364.
  4. Pan American Health Organization. Dengue: Symptoms, prevention and treatment. 2025.
  5. Barbé B, Verdonck K, El-Safi S, Khanal B, Teav S, Lilo Kalo JR, et al. Rapid diagnostic tests for neglected infectious diseases: Case study highlights need for customer awareness and postmarket surveillance. PLoS Negl Trop Dis. 2016; 10: e0004655.
  6. Matusali G, Colavita F, Carletti F, Lalle E, Bordi L, Vairo F, et al. Performance of rapid tests in the management of imported dengue fever cases in Lazio, Italy, 2014–2019. Int J Infect Dis. 2020; 99: 193–198.
  7. Yow KS, Aik J, Tan EYM, Ng LC, Lai YL. Rapid diagnostic tests for the detection of recent dengue infections: An evaluation of six kits on clinical specimens. PLoS One. 2021; 16: e0249602.
  8. Zammarchi L, Colao MG, Mantella A, Capobianco T, Mazzarelli G, Ciccone N, et al. Evaluation of a new rapid fluorescence immunoassay for the diagnosis of dengue and Zika virus infection. J Clin Virol. 2019; 112: 34–39.
  9. Liu LT, Chen CH, Tsai CY, Lin PC, Hsu MC, Huang BY, et al. Evaluation of rapid diagnostic tests to detect dengue virus infections in Taiwan. PLoS One. 2020; 15: e0239710.
  10. Chong ZL, Sekaran SD, Soe HJ, Peramalah D, Rampal S, Ng CW. Diagnostic accuracy and utility of three dengue diagnostic tests for acute dengue infection in Malaysia. BMC Infect Dis. 2020; 20: 210.
  11. Kikuti M, Cruz JS, Rodrigues MS, Tavares AS, Paploski IAD, Silva MMO, et al. Accuracy of the SD BIOLINE Dengue Duo for rapid point-of-care diagnosis of dengue. PLoS One. 2019; 14: e0213301.
  12. Santoso MS, Yohan B, Denis D, Hayati RF, Haryanto S, Trianty L, et al. Diagnostic accuracy of five brands of dengue virus nonstructural protein 1 antigen rapid diagnostic tests in Indonesia. Diagn Microbiol Infect Dis. 2020; 98: 115116.
  13. Zapata de la Cruz M, Guzmán Priego CG, Falcón Lezama J. Evaluation of the performance of rapid NS1 protein detection tests for dengue diagnosis [Master’s thesis]. Tabasco (Mexico): Universidad Juárez Autónoma de Tabasco; 2025.
  14. Zuroidah N, Tanzilia MF, Sunari IGAPE, Wrahatnala BJ, Haki FN, Aryati A, et al. Comparison of diagnostic tests for detection of dengue virus nonstructural protein 1 antigen using immunochromatography and fluorescence immunoassay methods. Res J Pharm Technol. 2022; 15: 1494–1498.
  15. Prabowo MH, Chatchen S, Rijiravanich P, Klamkam P, Chalermwatanachai T, Limkittikul K, et al. Clinical evaluation of a paper-based dengue NS1 rapid diagnostic test for febrile illness patients. Int J Infect Dis. 2021; 107: 271–277.
  16. Vickers IE, Harvey KM, Brown MG, Nelson K, DuCasse MB, Lindo JF. Performance of the SD BIOLINE Dengue DUO rapid immunochromatographic test during a dengue type 1 epidemic in Jamaica. J Biomed Sci. 2015; 22: 55.
  17. Ruchusatsawat K, Benjamungkalarak T, Phunikom N, Vateh H, Kowitdamrong E, Wongpiyabovorn J, et al. Performance comparison between fluorescent immunoassay and immunochromatography for rapid dengue detection. Sci Rep. 2022; 12: 17299.
  18. Nan X, Yang L, Cui Y. Lateral flow immunoassay for proteins. Clin Chim Acta. 2023; 544: 117337.
  19. Alcon S, Talarmin A, Debruyne M, Falconar A, Deubel V, Flamand M. Enzyme-linked immunosorbent assay specific to dengue virus type 1 nonstructural protein NS1 reveals antigen circulation during acute disease. J Clin Microbiol. 2002; 40: 376–381.
  20. Kabir MA, Zilouchian H, Younas MA, Asghar W. Dengue detection: Advances in diagnostic tools from conventional technology to point-of-care. Biosensors. 2021; 11: 206.
  21. Kok BH, Lim HT, Lim CP, Lai NS, Leow CY, Leow CH. Dengue virus infection: A review of pathogenesis, vaccines, diagnosis and therapy. Virus Res. 2022; 324: 1–17.
  22. Sirisena PDNN, Mahilkar S, Sharma C, Jain J, Sunil S. Concurrent dengue infections: Epidemiology and clinical implications. Indian J Med Res. 2021; 154: 669–679.
  23. Duyen HTL, Ngoc TV, Ha DT, Hang VTT, Kieu NTT, Young PR, et al. Kinetics of plasma viremia and soluble nonstructural protein 1 concentrations in dengue. J Infect Dis. 2011; 203: 1292–1300.
  24. Hang VT, Nguyet NM, Trung DT, Tricou V, Yoksan S, Dung NM, et al. Diagnostic accuracy of NS1 ELISA and lateral flow rapid tests for dengue. PLoS Negl Trop Dis. 2009; 3: e360.
  25. Blessmann J, Winkelmann Y, Keoviengkhone L, Sopraseuth V, Kann S, Hansen J, et al. Diagnostic and analytic performance of the SD BIOLINE Dengue Duo test in an endemic area. PLoS One. 2020; 15: e0230337.
  26. Ngim CF, Husain SMT, Hassan SS, Dhanoa A, Ahmad SAA, Mariapun J, et al. Rapid testing requires clinical evaluation for accurate dengue diagnosis. PLoS Negl Trop Dis. 2021; 15: e0009445.
  27. Tan LK, Wong WY, Yang HT, Huber RG, Bond P, Ng LC, et al. Flavivirus cross-reactivity to dengue nonstructural protein 1 antigen detection assays. 2020.
  28. Poltep K, Phadungsombat J, Kosoltanapiwat N, Hanboonkunupakarn B, Wiriyarat W, Suwanpakdee S, et al. Performance of the nonstructural protein 1 antigen rapid test for detection of all four dengue virus serotypes. Virol J. 2022; 19: 169.
  29. Matheus S, Boukhari R, Labeau B, Ernault V, Bremand L, Kazanji M, et al. Specificity of dengue NS1 antigen in differential diagnosis of dengue and Zika virus infection. Emerg Infect Dis. 2016; 22: 1691–1693.
  30. Lim JK, Chanthavanich P, Limkittikul K, Lee JS, Sirivichayakul C, Lee KS, et al. Clinical and epidemiologic characteristics associated with dengue fever in Thailand, 2011–2016. PLoS Negl Trop Dis. 2021; 15: e0009513.